UVCS MISSION LOG ---------------- Date: 25 May 2010 ----- Last revision: 26 July 2010 (RR) -------------- File Name: 2010.05.25.a.log ---------- Compiled by: R. Rosati ------------ UVCS Activity: Ramp down and turn off LyA detector -------------- Turn on and ramp up OVI detector Conduct OVI Stimulus Pulse Test (no exps taken due to problem with discriminator settings) Conduct OVI detector dark rate diagnostic Move LyA grating from 100000 to 67000 SOHO Activity: -------------- EOF Staff: L. Strachan: Lead Scientist ---------- R. Rosati: Commander Command file: 10.05.25.18:03:11.cmd.log Activity: --------- 1) LyA detector ramped down and turned off Sequence: Manual SCI_SPEC: Following completion of the first LyA component of the synoptic observation, the LyA MCP is ramped down and turned off. OBS_PROG: SCI_OBJ: OBJ: CMP_NAME: CMP_NUM: Duration: start: 2010.05.25.18:06 end: 2010.05.25.18:15 ONUMs = Instrument configuration: - Pointing: Sun center - Spacecraft Roll: 180 Degrees - Roll: -90 Degrees (51551) - Mirror/Occulter: 2.4 Ro 1.5 mm overocculting - LYA Slit: 50 um - OVI Slit: 100 um - LYA GRTPOS: 100000 - OVI GRTPOS: 210000 - LYA DETECTOR POWERED OFF - OVI DETECTOR POWERED OFF - LYA mask: Fits Files: 2) OVI detector turn on, ramped up, and OVI Stimulus Pulse Test conducted Sequence: Manual, TakeOVIexps.tcl 0 0 2 SCI_SPEC: After switching off LyA detector, the OVI detector is turned on a stimulus pulse test is completed After powering on OVI detector, discriminator settings were discovered to be incorrect. Discriminator settings were commanded to their nominal values. No stimilus pulse exposures were taken. OBS_PROG: SCI_OBJ: OBJ: CMP_NAME: CMP_NUM: Duration: start: 2010.05.25.18:17 end: 2010.05.25.18:47 ONUMs = Instrument configuration: - Pointing: Sun center - Spacecraft Roll: 180 Degrees - Roll: -90 Degrees (51551) - Mirror/Occulter: 2.4 Ro 1.5 mm overocculting - LYA Slit: 50 um - OVI Slit: 100 um - LYA GRTPOS: 100000 - OVI GRTPOS: 232100 - OVI MCP Voltage: 0 - OVI mask: Fits Files: 3) OVI Detector Dark Rate Diagnostic Sequence: OVI_dark_chk 26317 51551 SCI_SPEC: Determine Dark count rate for OVI detector. Instrument rolled to coronal hole, 5.0 Ro with 0.8 mm exposed mirror. Script engages OVI filter and then disengages it at end. OBS_PROG: Calibration SCI_OBJ: Calibration OBJ: Coronal Hole CMP_NAME: CMP_NUM: Duration: start: 2010.05.25.21:08 end: 2010.05.25.22:09 ONUMs = 26317 (12 exps) Instrument configuration: - Pointing: Sun center - Spacecraft Roll: 180 Degrees - Roll: -90 Degrees (51551) - Mirror/Occulter: 5.0 Ro 1.5 mm overocculting - LYA Slit: 50 um - OVI Slit: 100 um - LYA GRTPOS: 100000 - OVI GRTPOS: 210000 - OVI Filter Engaged - OVI MCP Voltage: 205 - OVI mask: OVI_chk_1xfull in slot 6 Fits Files: q10.05.25.21:31:41.ovi.dat